Instructions For Use - Otto Bock 1D35 Dynamic Motion Instructions For Use Manual

Hide thumbs Also See for 1D35 Dynamic Motion:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 7
► To avoid the risk of injury and product damage, only use the
product for a single patient.
► Comply with the product's field of application and do not expose
it to excessive strain (see page 7).
► Note the combination possibilities/combination exclusions in the
instructions for use of the products.
NOTICE!
Risk of product damage and limited functionality
► Do not expose the product to prohibited environmental condi­
tions.
► Check the product for damage if it has been exposed to prohib­
ited environmental conditions.
► Do not use the product if it is damaged or in a questionable con­
dition. Take suitable measures (e.g. cleaning, repair, replace­
ment, inspection by the manufacturer or a specialist workshop).
► To prevent mechanical damage, use caution when working with
the product.
► If you suspect the product is damaged, check it for proper func­
tion and readiness for use.
► Do not use the product if its functionality is restricted. Take suit­
able measures (e.g. cleaning, repair, replacement, inspection by
the manufacturer or a specialist workshop).
Signs of changes in or loss of functionality during use
Decreased forefoot resistance or changes in roll-over behaviour are
noticeable indications of loss of functionality.
4 Scope of delivery
Quantity
1
1
8
Designation

Instructions for use

Prosthetic foot
Spare parts/accessories (not included in the scope of deliv­
Designation
Connection cap
5 Preparing the product for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
NOTICE
Altering the prosthetic foot or footshell
Premature wear resulting from damage to the product
► Do not alter the prosthetic foot or footshell.
5.1 Bench Alignment
TT bench alignment
Bench alignment process
Required materials: 662M4 goniometer, 743S12 heel height
measuring apparatus, 743A80 50:50 gauge, alignment apparatus
(e.g. 743L200 L.A.S.A.R. Assembly or 743A200 PROS.A.
Assembly)
Perform the assembly and alignment of the prosthetic components
in the alignment apparatus according to the following specifica­
tions:
Sagittal plane
Heel height: Effective heel height (shoe heel height - sole
   
thickness of forefoot) + 5 mm
Exterior foot rotation: approx. 5°
   