6 Using the product
6.2.3.

Draft-Measurement

Calling up the function:
✓ A flue gas probe must be connected.
Performing the measurement:
-
Draft zeroing.
2. Position the flue gas probe in the hot spot (area of the highest
flue gas temperature).
The display showing the maximum measured flue gas
temperature
-
The reading is displayed.
Options:
[Options]
>
[Options]
>
is deleted.
[Options]
>
>
[Options]
form of a line graph.
>
[Options]
available during a measurement): The measurement view menu
is opened.
4.
60
The pressure socket of the instrument must be free (i.e.
unpressurized, not closed).
Do not measure for longer than 5 min, as the drift of the
pressure sensor means that the meas. views could be
outside the tolerance limits.
(FT
max) helps when positioning the probe.
→ Clipboard: Data are saved to the clipboard:
Delete
clipboard: Any data saved to the clipboard
→ Save: The meas. views are saved in a protocol.