Overview; Assigning Device Tag And Node Address; Configure Transmitter - Emerson Rosemount 5600 Series Reference Manual

Radar level transmitter with foundation fieldbus protocol
Hide thumbs Also See for Rosemount 5600 Series:
Table of Contents

Advertisement

Rosemount 5600 Series

Overview

ASSIGNING DEVICE TAG
AND NODE ADDRESS
CONFIGURE
TRANSMITTER
Table 4-1. Parameters that are
configurable for each antenna
type
4-2
Each F
fieldbus configuration tool or host device has a different
OUNDATION
way of displaying and performing configurations. Some will use Device
Descriptions (DD) and DD Methods to make configuration and displaying of
data consistent across host platforms. Since there is no requirement that a
configuration tool or host support these features, this section will describe
how to reconfigure the device manually. Appendix F: Operation with Delta V
shows the Delta V implementation of these common functions.
This section covers basic operation, software functionality, and basic
configuration procedures for the Rosemount 5600 Radar Transmitter with
F
fieldbus (Device Revision 1). For detailed information about
OUNDATION
F
fieldbus technology and the function blocks used in the
OUNDATION
Rosemount 5600 transmitter, refer to the F
(00809-0100-4783).
The Rosemount 5600 is shipped with a blank tag and a temporary address
(unless specifically ordered with both) to allow a host to automatically assign
an address and a tag. If the tag or address need to be changed, use the
features of the configuration tool. The tools basically do the following:
1. Change the address to a temporary address (248-251).
2. Change tag to new value.
3. Change address to new address.
When the device is at a temporary address, only the tag and address can be
changed or written to. The resource, transducer, and function blocks are all
disabled.
The wizard is used to do a standard configuration of the device. For further
information refer to the Rosemount 5600 manual (00809-0100-4024) and
Appendix B: Level Transducer Block of the Rosemount 5600 F
fieldbus manual. All settings that are made from this method can also be
made manually from the DD information, through the parameters listed below.
The following are configured by stepping through the
Configure Transmitter Wizard:
1. Choice of Antenna Type (ANTENNA_TYPE).
2. Based on antenna type choice, the different antenna related
configuration parameters will be available for configuration. See
Table 4-1 for Tank Connection Length (ANTENNA_TCL), and Antenna
Pipe Diameter (ANTENNA_PIPE_DIAMETER).
ANTENNA_TYPE
ANTENNA_TCL
User Defined
configurable
Cone
factory configured
Pipe
factory configured
Rod
factory configured
Process Seal
factory configured
Reference Manual
00809-0100-4025, Rev AA
fieldbus Block manual
OUNDATION
ANTENNA_PIPE_DIAM
GEOM_HOLD_OFF
configurable
factory configured
configurable
factory configured
factory configured
May 2003
OUNDATION
configurable
configurable
configurable
configurable
configurable

Advertisement

Table of Contents
loading

Table of Contents