Technical Data
Defibrillation
Operating modes:
Patient impedance:
Impulse shape:
Output power at:
Accuracy:
Impulse length:
Discharges:
Charge time:
ECG
Derivation:
Heart frequency:
Input:
Input resistance:
CMRR:
Input d.c. voltage:
Bandwidth:
Impedance measurement
Defibrillation:
Measurement frequency:
Operating Instructions
42
Asynchronous, external
23 – 200 Ohm
Biphasic, current regulated (CCD)
Patient impedance
25 Ohm
50 Ohm
75 Ohm
100 Ohm
125 Ohm
150 Ohm
175 Ohm
All data is subject to a tolerance of +/- 15%
Positive phase 11.25 ms, negative phase 3.75 ms
200 discharges at 20 °C with a new PRIMEDIC™ battery 6 at any energy of 360 J.
70 discharges at 20 °C with a new fully charged PRIMEDIC™ AkuPakat any energy
of 200 J.
50 discharges at 20 °C with a new fully charged PRIMEDIC™ AkuPakat any energy
of 360 J.
12 +/-3 seconds with a battery at 90% of the rated capacity
Einthoven II
30 – 300 min
-1
(Accuracy +/- 1/min, 1%)
Class BF, for 2-pin patient cable, defibrillation protected
> 5 MOhm @ 10 Hz
> 85 dB
± 0.5 V
0.5 – 40 Hz (- 3 dB) SR = 101 samples/s
23 ... 200 Ohm (accuracy +/- 20%)
30 kHz
HeartSave AED / AED-M, 21212 / GB / E02
1st. stage
2nd. stage
143 J
201 J
281 J
350 J
348 J
360 J
344 J
343 J
314 J
316 J
290 J
293 J
269 J
272 J
3rd. stage
277 J
360 J
360 J
343 J
317 J
293 J
272 J