5.10 Calling up/activating a profile

Certain parameter settings for the menu can be summarised into profiles. Saved
profiles can be called up as follows:
Note

5.11 Saving menu parameters in a profile

Certain parameter settings for the menu can be saved as profiles. The profiles Basic,
Profile 1, Profile 2, and Profile 3 are available. If you want to save parameters in a
profile or want to change a profile, proceed as follows:
If you want to change the configuration that your device starts up with when it is switched on,
you have to save your changed menu parameters in the profile "Basic".
Operating Instructions
30
Use the key
to change into the setup menu
Select your required profile using the keys   and confirm it with the key
The selected profile is active
Please note that your profile selection is only active until the device is
switched off. The devices always starts with the profile „Basic"
Use the key
to change into the setup menu
Change the required parameters from the various pages of the menu to suit your
needs.
Change to page 4 of the menu.
Use the key  to navigate to the entry < Profile selection > and confirm your
selection with the key
Use the keys   to select the required profile which is to be used to save the
menu parameters previously selected. Confirm this with the key
Navigate with the key  to the entry < PIN > and confirm your selection with the
key
Enter the current PIN as follows: Using the keys   you can increase or
decrease a digit. With the key
fourth digit, it jumps back to the menu
Change to the menu entry < Save to profile > and confirm your selection with the
key
On the right next to the cursor the entry < OK > should appear. This means the
profile is saved.
Now leave the menu by using the key  to navigate to the menu item < End
Setup > and confirm this with the key
HeartSave AED / AED-M, 21212 / GB / E02