Preparatory measures before (initial) start-up
5.5.2 Loading the PRIMEDIC
Basic / Comfort.
Please refer to the separate operating instructions for the charger Basic / Comfort.
5.5.3 Connecting up the PRIMEDIC
Fig. 15: PRIMEDIC
Procedure:
METRAX GmbH recommends that you keep the HeartSave unit, with the PRIMEDIC
Note
PowerLine inserted, continuously connected to the mains, so that the equipment self-
tests are carried out automatically.
5.6 Periodic device self-test
5.6.1 Self-test after switching on the PRIMEDIC
The device self-test is run automatically by switching the HeartSave on, or by inserting
the power module in the device to check all important functions and signal setups.
If the power module is replaced and the unit has first registered a fault, the large self-
test (FULL) is carried out automatically. Please proceed to follow the device
instructions.
Operating Instructions
26
TM
TM
PowerLine
Insert the PRIMEDIC
TM
PowerLine following the instructions in chapter 5.3.
Then plug the mains plug of the mains cable into a socket in the vicinity of the
patient.
The HeartSave carries out a self-test and is then ready for use.