6.7 Removing protective film from PRIMEDIC™ SavePads.
Figure 17: Removing protective film from PRIMEDIC™ SavePads.
The PRIMEDIC™ HeartSave AED will give a voice prompt for you to
apply the defibrillation electrodes to the patient.
When the device asks you to:
proceed as follows:
Make sure that with the device variant without a monitor, the red electrode symbol LEDs on the
membrane keyboard go out. With the device variant with a monitor, the sign on the display
must disappear.
Operating Instructions
34
1
Protective film
2
Electrode
< Position electrodes >
First, remove the protective film (1) from one electrode (2) and them immediately
place the electrode on the position you had ascertained previously. Refer to label
stuck on the back of the electrodes.
Proceed to remove the protective film from the second electrode and place it in its
position.
Smooth the electrodes onto the patient ensuring there are no air bubbles under
the electrodes!
Do not touch the floor, objects, clothing or other body parts
with the electrodes (once the protective film has been
removed). The conductive layer of gel on the electrodes
might be removed by this. A thinner layer of gel could reduce
the success of defibrillation or lead to burns of the skin under
the electrodes during defibrillation!
HeartSave AED / AED-M, 21212 / GB / E02