Operating the device and sequence of reanimation

6.13 Keeping the defibrillator ready for use

At the end of a reanimation, clean the device, replace the PRIMEDIC™ SavePads and
check or, if necessary, replace the power supply unit so that the PRIMEDIC™
HeartSave device is ready to use again as soon as possible. Charge the PRIMEDIC™
battery pack to ensure that sufficient energy is available for the device to be used
again.
If any malfunctions or noticeable problems occur, please contact your nearest
Note
service station.
6.14 Monitoring the patient with the PRIMEDIC
After successful defibrillation with the AED-M, the patient can be monitored by the
monitoring function using the SavePads electrodes already used whilst they are being
transferred to hospital. On the HeartSave AED-M you only have the lead II (Einthoven)
available. If in this situation, ventricular fibrillation is detected again, renewed
resuscitation can be carried out very quickly. For this purpose, the vehicle must be
stopped and the engine switched off to ensure the analysis is correct.
If you need to monitor the ECG of a patient in other situations, please use the 2-core
ECG patient cable.