Operating the device and sequence of reanimation

The parameter "Network filter" can only be changed temporarily while
Note
operating the device. After starting the device, the network filter is always
switched off initially.
6
Operating the device and sequence of
reanimation
The sequence of the reanimation is realised in the device according to the
recommended guidelines of the European Resuscitation Council (ERC): Resuscitation
2005, 67S1, S7—S23, or the American Heart Association (AHA), 2005: American Heart
Association (Guidelines for Cardiopulmonary Resuscitation and Emergency
Cardiovascular Care (CPR) 2005). Make sure that before using the device you have
completed suitable training.
On the HeartSave AED, the guideline for the sequence of operation is
Note
preconfigured in the factory. A software update is necessary for changing the
guidelines. If necessary, contact Metrax Service or your local dealer. After the
software update, the identification on the device must match the software
programmed.
On the HeartSave AED-M you can select the guideline used in the setup
Note
menu.
6.1 Switching the PRIMEDIC™ HeartSave on / off
6.1.1 Switching the PRIMEDIC™ HeartSave on
The device is automatically activated by removing its cover. If the device does not
switch on automatically, switch it on by pressing the on / off button. After this, all
buttons are unlocked, apart from the trigger button. Triggering defibrillation is only
activated once ventricular fibrillation (VF) has been ascertained.
Directly after switching it on, an internal self-test is carried out to check important
functions and signal devices. Standby is confirmed by a beep. It is important to ensure
that the loudspeaker is working.
6.1.2 Switching off the PRIMEDIC™ HeartSave
The PRIMEDIC™ HeartSave can be switched off in different ways:
By pressing the on/off button for approx. 3 seconds. A warning beep will sound
simultaneously.
By closing the cover of the device.
If the device does not recognise a signal for 10 minutes and if no button is
pressed, it switches off automatically.
HeartSave AED / AED-M, 21212 / GB / E02