Technical Data - Primedic HeartSave AED Operating Manual

Hide thumbs Also See for HeartSave AED:
Table of Contents

Advertisement

8
34
HeartSave AED / AED-M

Technical Data

Defibrillation
Operating Modes:
Pulse Form:
Output energy in case of:
Accuracy
Impulse length
Discharges:
Charging Time:
ECG
Lead:
Heart rate:
Input:
Input impedance:
CMRR:
Input direct voltage:
Bandwidth:
Impedance measurement
Impedance range:
Measure frequency:
Signal amplification
asynchronous, external in Auto Mode
biphasic, current-controlled (CCD)
Patient impedance
1. Step
25 Ohm
143 J
50 Ohm
281 J
75 Ohm
348 J
100 Ohm
344 J
125 Ohm
314 J
150 Ohm
290 J
175 Ohm
269 J
all figures are subject to a tolerance of +/- 15%)
positive phase 11,25 ms, negative phase 3,75 ms
– 70 discharges at 20 °C with a new fully charged
accumulator
– 100 discharges at 20 °C with a new battery
12 seconds with a full accumulator and / or with a full bat-
tery
12 seconds with an accumulator and / or a battery
with 90 percent of its nominal capacity
II
30 – 300 per minute
Class BF, for 2-pole patient cable, defibrillation proof
> 5 MOhm @ 10 Hz
> 85 dB
± 0,5 V
0,5 – 44 Hz (- 3 dB) SR = 101 Hz
0 ... 240 Ohm
1 Hz
The signal amplification of the ECG on the monitor
takes place automatically.
Operating Manual
2. Step
3. Step
201 J
277 J
350 J
360 J
368 J
360 J
343 J
343 J
316 J
317 J
293 J
293 J
272 J
272 J
20439 / 03.04 / 03 / engl.