Technical Data

10 Technical Data
Defibrillation
Operating modes:
Patient impedance:
Impulse shape:
Output power in AUTO mode at:
Accuracy:
Impulse length:
Discharges:
Charge time:
ECG
Lead:
Heart frequency:
Input:
Input resistance:
CMRR:
Input d.c. voltage:
Bandwidth:
Impedance measurement
Defibrillation:
Measurement frequency:
Operating instructions
HeartSave PAD 22289 / GB / C
Asynchronous, external
23 – 200 Ohm
Biphasic, current regulated (CCD)
st
Patient impedance
1
25 Ohm
143 J
50 Ohm
281 J
75 Ohm
348 J
100 Ohm
344 J
125 Ohm
314 J
150 Ohm
290 J
175 Ohm
269 J
All data is subject to a tolerance of +/-15 %
Positive phase 11.25 ms, negative phase 3.75 ms
50 discharges, at 20 °C, with a new battery
12 +/-3 seconds with a battery at 90% of the rated capacity
II
-1
30 – 300 min
(Accuracy +/- 1/min, 1 %)
Class BF, for 2-pin patient cable, defibrillation protected
> 5 MOhm @ 10 Hz
> 85 dB
± 0.5 V
0.5 – 44 Hz (- 3 dB) SR = 101 samples/s
23 ... 200 Ohm (accuracy +/-20 %)
30 kHz
nd
rd
stage
2
stage
3
201 J
277 J
350 J
360 J
360 J
360 J
343 J
343 J
316 J
317 J
293 J
293 J
272 J
272 J
stage
41