Cleaning, maintenance and dispatch

7
Cleaning, maintenance and dispatch

7.1 Cleaning

7.2 Servicing

Independently of the use of the device, we recommend regular visual inspection /
servicing of the PRIMEDIC™ HeartSave and its accessories by the user / service
technician on a regular basis, at least once a year.
Make sure that the housing, cable, PRIMEDIC™ SavePads and all the other accessories
are undamaged!
Servicing checklist:
Operating Instructions
40
Only clean the device when it is switched off and with the
electrodes unplugged. Before cleaning, remove the power
module from the device or if there is an optional mains
power supply, unplug it from the mains! Do not use any
dripping wet cloths to clean it. Never pour liquids over the
device and do not immerse it in any liquids!
Clean the PRIMEDIC™ HeartSave and all its accessories, such as the wall bracket,
with commercially available household cleaners.
Use a slightly damp, clean cloth.
Use ordinary wiping disinfectants to disinfect (e.g. Gigasept FF).
Check the expiry date of the PRIMEDIC™ SavePads.
Check the expiry date of the power module and if necessary, replace the parts
with original parts!
Check that the status display is showing "OK".
Check whether you can switch on the device.
Check whether the device carries out a self-test automatically after being switched
on!
Check whether the slot for the power supply is clean!
Check that the device is fully equipped!
Check that the device is labelled and that the label is legible.
HeartSave AED / AED-M, 21212 / GB / E02