Symbols In The Status Display - Primedic HeartSave AED Operating Instructions Manual

Hide thumbs Also See for HeartSave AED:
Table of Contents

Advertisement

Description of device
6 ECG display
7 Instructions
Fig. 7: PRIMEDIC

4.3 Symbols in the status display

Display
Operating Instructions
14
TM
SavePads AED
1
Cover of device
2
Utensils carrier with brief instructions
3
SavePads (defibrillation electrodes)
4
Disposable gloves (not sterile)
5
Disposable razor and respiration cloth
Meaning
Passed device self-test
Battery/battery pack capacity is sufficient
Battery/battery pack is discharged
Passed device self-test
Symbol also appears if no power module
has been inserted!
Symbol also appears if the use by date of
the power module has been exceeded.
Device may be defective
Device defective
HeartSave AED / AED-M, 21212 / GB / E02
Action to be taken
Device ready to use
Equipment can be used, it may be necessary to
charge or replace the battery / AkuPak.
Insert battery / AkuPak.
Check use by date, if necessary replace with
new ones.
Carry out large self-test by replacing the
battery / AkuPak or by switching the
equipment on again.
Have device repaired by authorised dealer.