Using The Equipment On Ships; Use Of Primedic™ Heartsave Units Together With A Primedic™ Battery On Ships In The Merchant Navy; Use Of Primedic™ Heartsave Units Together With A Primedic™ Akupak On Ships In The Merchant Navy - Primedic HeartSave AED Operating Instructions Manual

Hide thumbs Also See for HeartSave AED:
Table of Contents

Advertisement

Using the equipment on ships

16
Using the equipment on ships
16.1 Use of PRIMEDIC™ HeartSave units together with a
PRIMEDIC™ battery on ships in the merchant navy:
The use of the following PRIMEDIC™ HeartSave (M250)-units:
PRIMEDIC™ HeartSave PAD / AED / AED-M / HS6 / HS6-S
with the energy supply module
PRIMEDIC™ Battery 15VDC 2.8Ah LiMnO4
fulfils the EMC requirements of "Zone for the bridge and the open deck" asper the
"Guidelines for the Performance of Type Approvals" or "Test Requirements for
Electrical / Electronic Equipment and Systems" of the "Rules for Classification and
Construction", Book VI "Additional Rules and Guidelines" of "Germanischer Lloyd,"
2003.
16.2
Use of PRIMEDIC™ HeartSave units together with a
PRIMEDIC™ AkuPak on ships in the merchant navy:
PRIMEDIC™ HeartSave units with a PRIMEDIC™ battery, have been EMC checked for
compliance with "Test Requirements for Electrical / Electronic Equipment and Systems"
of "Germanischer Lloyd," 2003.
Because the "PRIMEDIC™ AkuPak 12VDC 1.2Ah NiCd", whilst supplying the
PRIMEDIC™ HeartSave (M250) units behave as the "PRIMEDIC™ battery", we can
transpose the experience gained from all EMC test results to the combinations of the
PRIMEDIC™ AkuPak with all PRIMEDIC™ HeartSave (M250) units.
These combinations fulfil, during supply mode, the requirements of "Zone for the bridge
and the open deck".
Recharging the PRIMEDIC™ AkuPak requires a PRIMEDIC™ ClipCharger as the charger
unit.
Recharging must be carried out in a dry room which is isolated by metallic screening
from "Zone for the bridge and the open deck".
The equipment combination always fulfils, during charging, or in parallel charging
mode, the EMC requirements of the "General Power Supply Zone".
HeartSave AED / AED-M, 21212 / GB / E02
55

Advertisement

Table of Contents
loading

This manual is also suitable for:

Heartsave aed-m