Technical Data - Primedic DefiMonitor XD Series Operating Manual

Table of Contents

Advertisement

Operating Manual
14
50

Technical Data

Defibrillation
Operating modes:
Synchronisation:
Impulse shape:
Output energy in AUTO Mode in case of:
Accuracy:
Impulse length:
Discharges:
Charging time:
ECG
Lead:
Heart rate:
Input:
Class CF, for 2-pole patient cable, defibrillation proof
Input impedance:
CMRR:
Input direct voltage:
Bandwidth:
NELLCOR®-pulseoximetry-module:
SpO
2
Indication range:
Calibration range:
Measurement precision:
Adults
New-born
For information about test procedures for the calibration ask manufacturer.
Wavelength:
Infrared:
Light density:
Operating mode:
Actualisation time: < 2 sec.
asynchronous, external in Auto/Manual Mode
SYNC only in Manual Mode
biphasic, current controlled (CCD)
Patient impedance
1. Step
25 Ohm
143 J
50 Ohm
281 J
75 Ohm
348 J
100 Ohm
344 J
125 Ohm
314 J
150 Ohm
290 J
175 Ohm
269 J
all figures are subject to a tolerance of +/- 15%)
positive Phase 11,25 ms, negative Phase 3,75 ms
70 discharges at 20 °C with a new fully charged accumula-
tor
12 +/-3 seconds with an accumulator with 90 percent of
its nominal capacity
I, II, III, aVr, aVf, Avr
30 – 300 per minute
> 5 MOhm @ 10 Hz
> 85 dB
± 0,5 V
0,5 – 44 Hz (- 3 dB) SR = 101 Hz
100 ... 0 %
100 ... 50 %
SpO2
100 ... 70 %
+/- 2 digits
69 ... 50 %
+/- 3 digits
50 ... 0 %
not specified
95 ... 70 %
+/- 3 digits
Red: 660 nm
920 nm
0,5 lumen/cm2
continuous
DefiMonitor XD
2. Step
3. Step
201 J
277 J
350 J
360 J
368 J
360 J
343 J
343 J
316 J
317 J
293 J
293 J
272 J
272 J
21324 / 09.05 / 02 / GB