Safety Notes For The User; Safety Notes For Protection Of The Patient - Primedic HeartSave AED Operating Instructions Manual

Hide thumbs Also See for HeartSave AED:
Table of Contents

Advertisement

Safety information

3.3 Safety notes for the user

3.4 Safety notes for protection of the patient

Operating Instructions
8
Only use the device on a patient if:
You have been authorised to do so as a result of
training!
You have ensured its operational safety before using it
and that it is in good condition.
The state of the patient requires or allows an application.
Before using the device, check that it is within the operating
temperature range. This applies for example, if the
defibrillator is stored in a rescue vehicle.
Do not use the device if it is defective (e. g. if the
defibrillation cable is damaged).
Only use the device on a patient if:
You have been authorised to do so as a result of training!
You have ensured its operational safety before using it and
that it is in good condition.
Before using the device, ensure that the device is within
the operating temperature range. This is important, for
example, when the device is stored in an emergency
services vehicle in winter. Do not use the device if it is
defective (e. g. if the defibrillation cable is damaged).
Only use the device with accessories, wearing parts and
disposable items which have proven to be completely safe
to use by being tested by a testing authority licensed to
test the device when equipped ready for use. These
conditions are fulfilled by all original PRIMEDIC™
accessories and wearing parts.
Use new, undamaged defibrillation electrodes before their
expiry date for every patient to avoid any possible burns to
the skin!
Only connect the adhesive electrodes to the PRIMEDIC™
HeartSave. Use of the electrode system with other devices
can lead to dangerous impedance currents for the patient!
Do not use the device in the immediate vicinity of other
sensitive equipment (e.g. measuring equipment that is
sensitive to magnetic fields) or strong sources of
interference which could affect the way the PRIMEDIC™
HeartSave works. Keep a sufficient distance away from
HeartSave AED / AED-M, 21212 / GB / E02
DANGER
DANGER

Advertisement

Table of Contents
loading

This manual is also suitable for:

Heartsave aed-m

Table of Contents