HeartSave AED / AED-M

General Safety Instructions

Do not use the device in presence of flammable substances (e.g. petroleum ether
etc.) or in an atmosphere enriched with oxygen or combustible gases / vapours!

Safety Instructions For You, The User

Use the device to treat a patient only if
– you have been authorised to use it by a training!
– you are convinced of the functional safety and the proper condition of
the device before using it!
– the condition of the patient requires and / or permits using the device.
Before using the device, check if the device is in the operating temperature
range. This, for example, applies if the defibrillator is kept in an ambulance.
Do not use the device in case of any defect (e.g. if the defibrillation cable is
damaged).
The infrared interface emits invisible optical radiation. The sending diode even
during continuous operation in the circuit used is 'Eye Save' according to IEC
60825-1 Class 1.

Safety Instructions for the Patient

Use the device to treat a patient only if
– you have been authorised to use it by a training!
– you are convinced of the functional safety and the proper condition of
the device before using it!
– the device clearly detects ventricular fibrillation (VF) and
tachycardia
ventricular
Before using the device, check if the device is in the operating temperature
range. This, for example, applies if the defibrillator is kept in an ambulance in
winter.
Do not use the device in case of any defect (e.g. if the defibrillation cable is
damaged).
Use the device only with accessories, wearing parts and expendable articles
whose suitability with regard to their safe use has been proved by a body
authorised to test the ready usability of devices. All original PRIMEDIC™ acces-
sories and wearing parts fulfil these conditions.
To avoid possible skin burns, use new and undamaged defibrillation electrodes
falling within the expiry date for each patient!
Connect the adhesive electrodes only to PRIMEDIC™ HeartSave AED. The use of
electrode systems with other devices can cause harmful leakage current to emit
on to the patient.
Do not operate the device in the immediate proximity of other sensitive equip-
ment (e.g. measuring instruments that react sensitively to magnetic fields) or
strong sources of interference that are likely to affect the functioning of
PRIMEDIC™ HeartSave AED. Maintain adequate distance from other therapeutic
and diagnostic energy sources (e.g. diathermy, high frequency surgery, magnetic
spin tomography). These devices can affect PRIMEDIC™ HeartSave AED and in-
terfere with its functioning.
(VT) and recommends a defibrillation!
Operating Manual
20439 / 03.04 / 03 / engl.