Safety information

3.2 General safety instructions

3.3 Safety notes for you, the user

Operating instructions
HeartSave PAD 22289 / GB / C
DANGER
The device may not be used in the vicinity of flammable
materials (e.g. cleaning solvents or similar) or in an
atmosphere enriched with oxygen or flammable
gases/vapours.
WARNING
Only use the device on a patient if
you have been authorised to do so as a result of training!
you have ensured its operational safety before you use it
and that it is in good condition.
the state of the patient requires or allows an application.
Before using it, check whether the device is in the operating
temperature range. This applies for example, if the
defibrillator is stored in a rescue vehicle.
Do not apply the device if it has been damaged or is
defective (e.g. if the defibrillator cable is damaged or the
housing has been damaged)
Do not touch the patient during defibrillation. Avoid any
contact between parts of the patient's body (such as bare
skin on head or legs), as well as conductible liquids (such as
gels, blood or salt solutions) and metallic objects around the
patient (such as bed frame or traction devices) which present
unintended paths for the defibrillation current.