Sorvall®
Centrifuges
Theory
of
Operation5.
RefrigerationCompressor
operation
is
similar
to
operation
in
the
HOLD
mode.
except
the
tem¬perature
control,
programmed
by
the
blue
set
pointer,
causes
ac
power
to
Howacross
the
closed
contacts
of
relay
K205.The
timing
circuit
is
programmed
for
a
minimum
of
40
seconds
OFF
cycle
and
anunlimited
time
ON
cycle.
This
circuit
is
activated
by
the
presence
of
a
signal
fromthe
tachometer
generator
TG301.6.
Overtemperature
Detection
SystemThe
overtemperature
detector
consists
of
a
signal
comparator,
control
amplifier
andrelay
K4.
If
the
temperature
rises
above
the
red
set
pointer,
a
temperature
signal
isgenerated
by
the
comparator
to
open
the
overtemperature
switch
triac
CR20
anddeenergize
run-hold
relay
K205.
placing
the
centrifuge
in
the
HOLD
mode.7.
Brush
Warning
SystemThe
brush
warning
system
consists
of
circuit
breakers
CB
1 and
CB2
and
two
nor¬mally
open
brush
switches
in
the
drive
motor.
As
the
brush
wears,
the
switchslowly
closes,
until
contact
is
made,
connecting
ac
motor
voltage
to
its
correspond¬ing
circuit
breaker
through
a
250-ohm
resistor.
Voltage
above
12
to
20
volts
tripsthe
breaker.
The
tripped
circuit
breaker
opens
the
115
Vac
connection
between
run-hold
relay
K205
and
ON
switch
Sill
which
energizes
thermal
relay
K207.
Thermalrelay
K207
switches
ON
and
OFF
providing
an
intermittent
voltage
to
BRUSHESindicator
lamp
DS103.
When
circuit
breakers
trip,
remaining
brush
life
is
approxi¬mately
10
hours.
Once
the
run
is
completed,
the
centrifuge
will
not
restart
until
thecircuit
breakers
are
reset.8.
Motor
BrakingThe
motor
brakes
only
as
the
centrifuge
enters
or
is
in
the
HOLD
mode.
However,the
braking
function
can
be
set
at
any
time.
Pressing
BRAKE
switch
S109A
ener¬gizes
brake
relay
K208
to
apply
+5
Vdc
to
contact
9
of
motor
relay
K202.
PressingOFF
switch
Sl
10
deenergizes
motor
relay
K202
to
apply
the
+5
Vdc
to
the
Junctionof
resistors
R201
and
R202
to
brake
the
motor.
BRAKE
lamp
DS109
is
energized
by30
Vac
through
S
109B.3-3.
RC-5B
Theory
of
Operation
(figures
3-11
through3-24)The
theory
of
operation
for
the
RC-5B
centrifuge
Is
similar
to
that
of
the
RC-5,
with
thefollowing
exceptions:a.
Operational
ModesThe
RC-5B
is
capable
of
operating
in
the
HOLD
or
TIMED
modes.
Under
normal
condi¬tions,
the
compressor
will
continue
to
operate
in
the
STOP
mode.
The
RC-5B
will
alsooperate
In
the
SLOW
mode
when
equipped
with
the
optional
automatic
rate
controller.3-5