Maintenance
&
Repair
Sorvall®
Centrifuges2.
Installationa.
Mount
the
microswitch
to
the
centrifuge
with
the
two
screws.b.
Reconnect
the
terminating
wires.c.
Replace
the
deck
and
console.d.
Reinstall
the
four
cabinet
panels.m.
Door
Solenoid
L301
(RC-5B)1.
Removala.
Set
main
circuit
breaker
to
OFF.b.
Unplug
the
power
cord
from
socket.c.
Remove
the
deck
and
console
as
described
in
paragraph
6-12.
steps
3
through8.d.
Tag,
then
remove
all
terminating
wires.e.
Remove
the
four
screws
holding
the
solenoid
to
the
centrifuge,
disconnect
thespring
to
the
pivot
arm,
then
remove
the
solenoid.2.
Installationa.
Mount
the
solenoid
to
the
centrifuge
with
the
four
screws
and
connect
thespring
to
the
pivot
arm.b.
Reconnect
all
terminating
wires.c.
Replace
the
deck
and
console.d.
Reinstall
the
four
cabinet
panels.n.
Tachometer
Indicator
Ml
011.
Removala.
Set
main
circuit
breaker
to
OFF.b.
Unplug
the
power
cord
from
socket.c.
Remove
console
rear
panel.d.
Tag
(+)
and
(—)
wires
on
M101.e.
Remove
both
terminal
nuts
and
terminated
(+)
and
(—)
wires.6-36

Hide quick links:

Advertisement

loading

This manual is also suitable for: