Calibration
Procedures
Sorvall®
Centrifugesi.
Power
Supply,
Temperature
Control:
+5
V
(R7)
and
+10
V
(R4)This
procedure
adjusts
the
temperature
control
power
supply
to
+5
V
and
+10
V
outputvoltage.1.
Set
the
circuit
breaker
to
OFF.2.
Connect
the
dc
voltmeter
between
TP-1
(-)
and
TP-7
(+).3.
Place
the
system
in
the
HOLD
mode.4.
Adjust
potentiometer
R4
for
+10
Vdc5.
Set
the
main
circuit
breaker
to
OFF.6.
Connect
the
dc
voltmeter
between
TP-1
(-)
and
TP-9
(+),
observing
proper
polarity
ofthe
test
leads.7.
Place
the
system
in
the
HOLD
mode.8.
Adjust
potentiometer
R7
for
+5
Vdc
indication
on
the
voltmeter.9.
Set
the
main
circuit
breaker
to
OFF.10.
Remove
the
voltmeter.j.
Temperature
Indicator
M102
Deflection
Adjust
-20°CPotentiometer
(R41)
and
+40°C
Potentiometer
(R40)This
procedure
adjusts
M102
for
zero
and
full
scale
indications.1.
Set
the
circuit
breaker
to
OFF.2.
Place
the
centrifuge
system
in
the
HOLD
mode.3.
Adjust
potentiometer
R40
fully
counterclockwise.4.
Referring
to
figure
7-2,
place
the
106.15
ohm
precision
resistor
in
the
circuit.5.
Adjust
potentiometer
R41
for
a -20°C
indication.
Remove
the
106.15
ohm
resistor.6.
Referring
to
figure
7-2,
place
the
149.79
ohm
precision
resistor
in
the
circuit.7.
Adjust
potentiometer
R40
for
a
+40°C
indication
.8.
Repeat
steps
4
through
7
until
-20°C
and
+40°C
indications
are
obtained.9.
Set
the
main
circuit
breaker
to
OFF.10.
Remove
the
precision
resistor
and
reconnect
the
original
circuit
as
shown
in
figure7-2.7-6