Du Pont SORVALL RC-5 Service Manual page 185

Automatic superspeed refrigerated centrifuges

Advertisement

Calibration
Procedures
Sorvall®
Centrifuges7-5.
Test
Equipment
RequiredVolt/Ohmeter:
Simpson
260,
Triplet
(FET-310)
or
equivalent.Precision
Resistor.
149.79
ohm.
0.1%.
Precision
Resistor.
106.15
ohm.
0.1%.Ammeter
(clamp-on
type):
Triplet
Model
10
or
equivalent.Portable
TachometerThermometer:
McAlaster
No.
39682
(-35°C
to
-50°C)
or
equivalent.7-6.
Calibration
Proceduresa.
Speed
Indicator
Ml
01
Zero
AdjustThe
meter
pointer
should
indicate
zero
at
zero
rotor
rpm.1.
Set
the
main
circuit
breaker
to
OFF.2.
While
gently
tapping
face
of
indicator
Ml
01,
turn
front
access
zero
adjust
clock¬wise.
until
pointer
aligns
with
zero
Indication
during
the
upswing.b.
Temperature
Indicator
M102
Zero
AdjustThe
temperature
pointer
(black)
should
indicate
-20°C
when
the
main
circuit
breaker
isat
OFF.1.
Set
the
main
circuit
breaker
to
OFF.2.
While
gently
tapping
the
face
of
M102,
turn
the
front
access
zero
adjust
until
thetemperature
pointer
aligns
with
-20°C
indication.c.
Speed
Control
Potentiometer
R106This
is
a
mechanical
alignment
of
SPEED
selection
for
reference
only.1.
Turn
the
SPEED
control
knob
fully
counterclockwise,
then
loosen
the
two
knob
set-screws.2.
Align
the
knob
0
(zero)
indication
with
the
SPEED
indication
and
secure
the
twosetscrews.d.
Source
PowerThis
is
a
check
of
the
main
power
to
assure
proper
operation
of
centrifuge.1.
Check
the
source
voltage.7-2