Du Pont SORVALL RC-5 Service Manual page 108

Automatic superspeed refrigerated centrifuges

Advertisement

Sorvall®
CentrifugesDrive
System
CircuitTable
4-10.
Troubleshooting
Chart,
Automatic
RateController
Sheet
4
of
4PROBLEM:
System
remains
in slow
start;does
not
transfer
to RUN.
Operatesnormally
in RUN
mode.System
Conditions:1.
SLOW
switch
ON.2.
"SPEED"
control
set
to
more
than
1000
rev/min.3.
START
switch
ON
(START
lamp
lit).4.
Door
closed.5.
RATE
control
at
maximum
setting.Check
tachometer
voltage
acrossR25.
At
1000
rev/min.
voltageshould
be 0.7
V ac which
shoulddouble
for
each
addtional
1000rev/min.
Is proper
voltagepresent?YESNORemove
power.
Check
continuitybetween
P109-6
and
terminal1021
on
PC-50195,
and
betweenP109-3
and
terminal
1021
onPC-50195.
If open,
repair
orreplace
as
required.At
some
"speed"
above
1000
rev/min.check
for
0V
across
CR11.
Isvoltage
correct?0V12VdcReplace
PC-50155.Check
for
115
V ac
betweenterminal
701
on
PC-50195and
gnd.
Is voltage
present?0V115
V
dcReplace
K206.Remove
K206.
Check
for115
V ac
between
pin
A
onrelay
socket
and
gnd.
Isvoltage
present?Remove
power
and
checkcontinuity
between
pin
A
ofK206
socket
and
terminal
1007on
PC-50195.
If open,
repairor
replace
as
required.8/79RCS/1524-27