Sign In
Upload
Download
Add to my manuals
Delete from my manuals
Share
URL of this page:
HTML Link:
Bookmark this page
Add
Manual will be automatically added to "My Manuals"
Print this page
×
Bookmark added
×
Added to my manuals
Manuals
Brands
Du Pont Manuals
Laboratory Equipment
SORVALL RC-5
Service manual
Du Pont SORVALL RC-5 Service Manual page 191
Automatic superspeed refrigerated centrifuges
Hide thumbs
1
2
3
4
5
6
Table Of Contents
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
214
215
216
217
218
219
220
221
222
223
224
225
226
227
228
229
230
231
232
233
234
235
236
237
238
239
240
241
242
243
244
245
page
of
245
Go
/
245
Bookmarks
Advertisement
Calibration
Procedures
Sorvall®
Centrifuges6.
Repeat
steps
2
through
5
until
-3°C
set-pointer
is
5°C
below
the
temperature
re¬corded
in
step
4.1.
Temperature
Control
Set-Points
Potentiometer
(R42
and
R43)This
procedure
establishes
the
RUN
and
HOLD
mode
compressor
set
points.1.
Set
the
main
circuit
breaker
to
ON.2.
Install
an
SS-34
rotor
and
latch
the
chamber
door.3.
Set
the
RUN
mode
set
point
to
5°C
below
indicated
temperature.4.
Set
the
HOLD
mode
set
point
to
5°C
above
Indicated
temperature.5.
Connect
the
dc
voltmeter
to
TP-1
(-)
and
the
collector
of
transistor
Q15
(+).
Meterindication
should
be
greater
than
4.5
Vdc.6.
Move
the
HOLD
set-pointer
toward
the
back
temperature
indicator
pointer
until
theHOLD
pointer
is
exactly
over
the
black
temperature
indicator
pointer.
The
voltageshould
switch
to
lower
than
1.0
Vdc.7.
If
the
voltage
change
occurs
above
or
below
the
set-pointer,
align
the
set-pointerwith
the
black
indicator
and
adjust
potentiometer
R42
accordingly.8.
Set
the
SPEED
control
to
1000
rpm.9.
Be
sure
the
rotor
is
secured
and
the
chamber
door
latched.10.
Set
the
HOLD
mode
set
point
to
5°C
above
indicated
temperature.11.
Set
the
RUN
mode
set
point
to
5°C
below
Indicated
temperature.12.
Place
the
system
in
the
RUN
mode.13.
The
meter
indication
should
be
less
than
1.0
V
dc.14.
Move
the
RUN
set
point
towards
the
black
temperature
indicator
pointer
until
thepointer
is
completely
aligned
with
the
black
temperature
Indicator
pointer.
Thevoltage
should
switch
to
greater
than
+4.5
Vdc.15.
If
the
voltage
change
occurs
above
or
below
the
set
point,
align
the
set
point
with
theblack
indicator,
then
adjust
potentiometer
R43
accordingly.m.
Overtemperature
Detection
Set-Pointer
(R58)This
procedure
adjusts
the
overtemperature
set
point
to
inhibit,
or
disable
the
operationof,
the
RUN
mode.1.
Set
the
main
circuit
breaker
to
ON.7-8
Previous
Page
Next
Page
1
...
190
191
192
193
Show Quick Links
Quick Links:
Table of Contents
Hide quick links:
Permanently
Temporary
Cancel
Advertisement
Related Manuals for Du Pont SORVALL RC-5
Laboratory Equipment Du Pont SORVALL RT6000B Operating Instructions Manual
Tabletop centrifuges (28 pages)
Laboratory Equipment Du Pont Sorvall RC-5B Plus Service Manual
(102 pages)
Laboratory Equipment Du Pont SORVALL SUPRASPEED RC-28S Operating Instructions Manual
(71 pages)
Related Products for Du Pont SORVALL RC-5
Du Pont SORVALL RT6000B
Du Pont Sorvall RC-5B Plus
Du Pont SORVALL SUPRASPEED RC-28S
Du Pont SORVALL T6000B
This manual is also suitable for:
Sorvall rc-5b
Print
Rename the bookmark
Delete bookmark?
Delete from my manuals?
Login
Sign In
OR
Sign in with Facebook
Sign in with Google
Upload manual
Upload from disk
Upload from URL