Maintenance
&
RepairSon/all®
CentrifugesFigure
6-51.Removal
of
Door
MicroswitchS301Figure
6-52.Removing
Mounting
Hardwarefor
Door
Microswitchb.
Connect
wires
to
the
microswitch.c.
Replace
the
top
deck
and
console.d.
Reinstall
the
four
cabinet
panels.k.
Door
Solenoid
L301
(RC-5)1.
Removala.
Set
main
circuit
breaker
to
OFF.b.
Unplug
the
power
cord
from
socket.c.
Remove
the
deck
and
console
as
described
in
paragraph
6-12,
steps
3
through8.d.
Using
an
Alien
wrench
(9/64
in),
remove
the
two
Alien
head
screws
holding
thesolenoid
assembly
to
the
centrifuge.e.
Straighten
the
set
portion
of
the
wire
used
to
manually
operate
the
solenoid
toremove
the
right
angle
bend.f.
Disconnect
the
Amphenol
plug.
the
remove
the
assembly,
being
careful
not
todamage
the
wire
used
to
manually
operate
the
solenoid.6-34