Download Print this page

Du Pont SORVALL RC-5 Service Manual page 196

Automatic superspeed refrigerated centrifuges

Advertisement

Son/all®
Centrifuges
Calibration
ProceduresTable
7-2.
Test
Point
Description
forP.C.
Board,
PN
50059
(Continued)NOTERefrigeration
system
will
also
be
off
for
two,
minimum
of
40
seconds
onrun
cycle.
Hold
mode
compressor
is
on
no
longer
than
15
seconds
andmust
be
off
for
two
minutes.TP-7
-
TP-1
10
Vdc
Temperature
Control
Power
Supply.
Lights
lamps
in
PhotoCell
and
power
for
temperature
meter
bridge
circuit.TP-8
-
TP-1
Controls
compressor
0 Volts
Compressor
on,
5
Vdc
Compressor
off.TP-9
-
TP-1
5
Vdc
power
to
temperature
control
logic.Table
7-3.
Potentiometer
Adjustment
forP.C.
Board.
PN
50059R4
TP-1
-
TP-7.
10
Vdc
Temperature
Control
Pilot
and
Bridge
Circuit.NOTEAdjust
R4
before
adjusting
R7.R7
TP-1
-
TP-9.
5
Vdc
Power
to
Temperature
Control
Logic.R9
Speed
Control
High
rpm
Setting
20
000
rpm.R18
Speed
Control
Low
rpm
Setting
1000
rpm.*R40
Temperature
Control
Indicating
Needle
+40
Set
Point.*R41
Temperature
Control
Indicating
Needle
-20
Set
Point.NOTEUse
temperature
calibrating
test
set
when
adjusting
R40
and
R41.*R42
Temperature
Control
Orange
Pointer
(hold).
Adjust
to
turnCompressor
on
when
Temperature
Indicating
Needle
meets
HoldTemperature
Set
Point.7-13