Download Print this page

Du Pont SORVALL RC-5 Service Manual page 152

Automatic superspeed refrigerated centrifuges

Advertisement

Maintenance
&
Repair
Sorvall®
Centrifugesb.
Installation1.
Install
the
component.2.
Reconnect
the
wires.3.
Install
the
capacitor
box
cover.6-10.
Condenser
Fan
Replacementa.
Removal1.
Set
main
circuit
breaker
to
OFF.2.
Unplug
the
power
cord
from
socket.3.
Remove
the
rear
cabinet
panel.4.
Remove
the
four
bolts
securing
the
fan
motor
and
bracket
to
the
condensing
assem¬bly
deck.5.
Move
the
motor
and
bracket
away
from
the
evaporator
fins.6.
Mark
orientation
of
fan
on
the
motor
shaft.
Loosen
the
setscrew
on
the
fan
and
re¬move
the
fan
from
the
motor
shaft.b.
Installation1.
Vacuum
all
dust
accumulation
from
the
evaporator
fins.2.
If
a
used
fan
is
to
be
installed,
check
the
blades
and
adjust
them
in
accordance
withFigure
6-17.3.
Orient
the
fan
on
the
motor
shaft
(see
removal
step
6)
and
secure
the
fan
by
tighten¬ing
the
setscrew
on
the
flat
surface
of
the
motor
shaft,
making
sure
that
the
sur¬faces
of
the
motor
shaft
and
fan
hub
are
flush.4.
Secure
the
motor
bracket
to
the
condenser
assembly
deck
with
the
four
bolts.5.
Reinstall
the
rear
cabinet
panel.6-14

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Sorvall rc-5b