Du Pont SORVALL RC-5 Service Manual page 181

Automatic superspeed refrigerated centrifuges

Advertisement

Sorvall®
Centrifuges
Maintenance
&
Repaire.
Reinstallation
of
Components1.
Reinstall
counter
spring
adjuster
and
adjust
so
that
door
will
not
strike
the
latchwhen
allowed
to
fall
from
the
fully
open
position.2.
Refit
the
solenoid.
Shim
as
necessary
with
special
rectangular
0.8
mm
(1/32
in)shims.
The
interlock
pin
must
travel
free.
Refit
and
adjust
the
microswitch.3.
Reinstall
the
deck
and
console
and
check
that
neither
hinge
nor
latch
interfereswith
the
lip
of
the
deck
cover
in
any
position.6-16.
Refrigeration
Maintenancea.
Compressor
Resistance
ValuesTecumseh
compressor
motor
resistance
values
are
measured
across
compressor
termi¬nal
leads.
Approximate
resistance
of
motor
windings
at
ambient
temperature
of
25°C(77°F)
is
given
in
Table
6-1.
Resistance
of
both
windings
is
measured
across
compressorterminals
R
and
S
to
common
C.b.
Compressor
BurnoutIf
a
compressor
failure
has
been
determined,
it
will
be
necessary
to
replace
the
completecondensing
assembly.
In
this
case,
a
determination
must
be
made
of
the
type
of
burn¬out
that
has
occurred
so
that
the
evaporator
liner
can
be
cleaned
accordingly.Two
types
of
burnouts
can
occur:
a
standing
burnout
or
a
running
burnout.
These
canbe
determined
by
smelling
and
visually
examining
the
refrigerant.1.
Open
a
service
valve
while
holding
a
white
cloth
or
white
tissue
over
the
opening
tocatch
a
few
drops
of
oil.2.
Sniff
the
refrigerant;
an
acrid
odor
indicates
a
burnout.3.
Examine
the
refrigerant.
If
the
oil
is
clean,
a
standing
burnout
is
indicated
and
thecontamination
of
the
evaporator
liner
will
not
be
extensive.Table
6-1.
Tecumseh
Compressor
Motor
Resistance
ValuesTecumseh
ModelAH2511JAH2466AJE150JB100ResistanceSTART1-53-63-66-9Range
(Ohms)BOTH2-5}3-6}4-7)7-10}Terminal
StripLocationcS^R000C
S
R6-43