Download Print this page

Du Pont SORVALL RC-5 Service Manual page 173

Automatic superspeed refrigerated centrifuges

Advertisement

Son/all®
CentrifugesMaintenance
&
Repair2.
Installationa.
Install
the
solenoid
on
the
centrifuge
then
fasten
it
with
the
two
Alien
headscrews.b.
Connect
the
Amphenol
plug.c.
Bend
the
manual
operate
wire
into
a
right
angle
bend
at
the
front
of
the
centri¬fuge.d.
Replace
the
deck
and
console.e.
Reinstall
the
four
cabinet
panels.1.
Door
Release
Microswitch
S303
(RC-5B)1.
Removala.
Set
main
circuit
breaker
to
OFF.b.
Unplug
the
power
cord
from
socket.c.
Remove
the
top
deck
and
console
as
described
in
paragraph
6-12,
steps
3through
8.d.
Tag,
then
disconnect
all
terminating
wires.e.
Remove
two
screws
holding
the
microswitch
to
the
centrifuge.Figure
6-53.Removing
Hardware
MountingDoor
SolenoidFigure
6-54.Disconnecting
Wires
Terminatingat
Door
Microswitch
S3036-35