Du Pont SORVALL RC-5 Service Manual page 118

Automatic superspeed refrigerated centrifuges

Advertisement

Refrigeration
SystemSon/all®
CentrifugesRTDRELAYSELECTIONCOMPARATOR
REQUESTHOLD
TIMING(ENFORCED)15
SEC
ON120
SEC
OFFRUN
TIMING40
SEC
OFF(ENFORCED)NO
ON
LIMIT7/79RCS/44Figure
5-4.
Temperature
Control
Circuit,
Block
DiagramThe
selection
between
the
red
and
blue
set
points
is
made
by
the
RUN-HOLD
relay
K205and
relays
K2
and
K3.
Relay
K205
and
K2
are
energized
when
the
centrifuge
is
in
theRUN
mode,
and
refrigeration
is
controlled
by
the
blue
pointer.
K205
is
unenergized
andK3
is
energized
when
the
centrifuge
is
In
the
HOLD
mode,
and
refrigeration
is
controlledby
the
red
pointer.The
refrigeration
signal
is
applied
to
the
differential
amplifier
and
compared
against
areference
voltage.
When
the
refrigeration
signal
is
higher
than
the
reference
voltage,
K55-6