Du Pont SORVALL RC-5 Service Manual page 176

Automatic superspeed refrigerated centrifuges

Advertisement

Maintenance
&
RepairSon/all®
Centrifugesg.
Calibrate
in
accordance
with
table
7-1.h.
Replace
the
console
rear
panel.o.
Temperature
Indicator
M1021.
Removala.
Set
main
circuit
breaker
to
OFF.b.
Unplug
the
power
cord
from
socket.c.
Remove
the
console
rear
panel.d.
On
the
front
panel,
loosen
the
setpoint
knob
setscrews
and
remove
both
knobs.e.
On
the
rear
of
the
console,
disconnect
connector
J
105.f.
Tag
the
(+)
and
(—)
terminated
wires,
then
remove
the
terminal
nuts
and
theterminated
wires.g.
Remove
the
lower
left
nut,
lockwasher,
and
Hat
washer,
and
remove
the
cableclamp,
if
present.h.
Remove
four
outer
mounting
screws
and
lockwashers,
then
remove
M
102
fromthe
console.1.
Remove
remaining
three
nuts
and
lockwashers,
and
remove
brackets
fromM102.Figure
6-57.
Disconnecting
Plug
and
Wires
fromTemperature
Indicator
Ml
026-38