Download Print this page

Du Pont SORVALL RC-5 Service Manual page 141

Automatic superspeed refrigerated centrifuges

Advertisement

Sorvall*
CentrifugesMaintenance
&
RepairFigure
6-1.Cabinet
Panel
Screw
RemovalFigure
6-2.Lifting
the
Panel
OffCAUTIONDo
not
operate
the
centrifuge
by
holding
circuit
breaker
in
the
ON
posi¬tion
before
removing
wire
connection
from
the
interlock;
to
do
so
willdamage
the
circuit
breaker.6-4.
Drive
Motor
Replacementa.
Removal1.
Set
main
circuit
breaker
to
OFF.2.
Unplug
the
power
cord
from
socket.3.
Remove
the
front
cabinet
panel.4.
Disconnect
the
motor
plug
from
inner
left
centrifuge
wall.
The
plug
is
flush
againstthe
cushion
under
the
chamber
part
of
the
centrifuge
and
held
there
by
a
springwhich
clips
into
the
slot
of
the
screw
on
top
of
the
plug.
To
remove
the
plug,
onefinger
must
be
placed
between
the
plug
and
the
cushion
to
release
the
spring
whileusing
the
other
hand
to
pull
the
plug
out
of
the
centrifuge
wall.6-3