Du Pont SORVALL RC-5 Service Manual page 166

Automatic superspeed refrigerated centrifuges

Advertisement

SorvalP
CentrifugesMaintenance
&
Repaird.
Circuit
Breaker
CB301
(RC-5)1.
Removala.
Set
main
circuit
breaker
to
OFF.b.
Unplug
the
power
cord
from
socket.c.
Remove
the
cabinet
front
panel.d.
Remove
three
screws,
then
extend
the
switch
plate
assembly.e.
Loosen
and
remove
the
key
switch
inner
locknut
behind
the
switch
plate
as¬sembly.f.
Remove
the
outer
locknut.
then
remove
the
keyswitch.g.
Remove
the
surface
plate.TB
201(RED)(BLK)TRANSFERFROMTB
201
-9BUCK/BOOSTTRANSFORMER(WHT)(463>.COMPRESSORCABLETRANSFER—FROM-TB
201
-87/79RCS/95Figure
6-47.
Transformer
Wiring/or
RC-5,
208
V,
60
Hz6-28