Download Print this page

Du Pont SORVALL RC-5 Service Manual page 127

Automatic superspeed refrigerated centrifuges

Advertisement

P/N: 50589-3
INSTRUCTIONS
RC-5B / RC-5C Preventative Maintenance Kit
(Kit Catalog No. 12010)
3. Inspect the rotor lid locking stud.
4. Clean and inspect any Sorvall® RX rotor covers. If lid does not fit on
the rotor smoothly replace the pertinent components.
5. Inspect "O"-rings (if applicable).
6. Record the serial numbers of all rotors inspected and any problems
relating to the rotors.
Clean the centrifuge cabenet and drive spindle:
The entire centrifuge must be kept clean. Clean the enameled cabinet
panels weekly with a commercial wax based cleaning product. Clean the
top deck surface and door with a household detergent and water. Use a
brush to remove ground in dirt. Immediately afterwards, rinse thoroughly
with water, especially if using solvent based or chlorinated cleaners for
removal of difficult stains. The drive spindle must be kept clean so the
rotor does not stick to it. Wipe it with a soft cloth each time before the ro-
tor is installed, warm water may be used for cleaning when the chamber is
defrosted. A clean undamaged surface is essential for proper rotor installa-
tion and removal.
for the
Cleaning Rotor Bearing Surfaces
Figure 1
Page 3