INSTRUCTIONS
RC-5B / RC-5C Preventative Maintenance Kit
(Kit Catalog No. 12010)
Incandesent light bulb replacement RC-5B:
Replace any light bulbs that may be out on the control console.
1. Remove the plastic indicator lense cover by pulling the cover straight
up from the control console (items 1 through 11 on the drawing shown
in figure #2.
2. Using a small needle nose pliers pull the light bulb from the socket.
3. Replace the bulb using one of the bulbs that are supplied with the PM
kit.
4. Replace the indicator lense.
Incandesent light bulb replacement RC-5C:
1. Using a Philips screwdriver remove the (6) screws that hold the con-
trol panel from the control console.
2. Lay the control panel (face down) on the centrifuge deck.
3. Remove the appropriate light bulb by loosening the bulb clip using a
small flat blade screwdriver. Rotate the clip 90 ° and remove the af-
fected bulb. Replace the bulb with a bulb that is included with the PM
Kit.
4. Rotate the clip 90º covering the bulb and tighten the clip.
5. Re-install the control panel into the console.