Download Print this page

Du Pont SORVALL RC-5 Service Manual page 153

Automatic superspeed refrigerated centrifuges

Advertisement

Sorvall®
CentrifugesMaintenance
&
RepairFigure
6-16.Removing
Fan.
Motor
BoltNOTES
and DIMENSIONSA
- 2-1/4
±
1/32B
- 1-11/32
REFC
- 29/32
±
1/32D
- PITCH ANGLE
23°E
- MEASUREMENT
TAKENFLUSH
AT
BLADESUPPORT
AND
HUBF
- LEADING
OR TRAILINGEDGES
OF BLADESMAY
BE OUT OF LINE1
/32
TO
0.050Figure
6-17.Compressor
Fan
BladeSpecifications6-11.
Condenser
Fan
Motor
Replacementa.
Removal1.
Set
main
circuit
breaker
to
OFF.2.
Unplug
the
power
cord
from
socket.3.
Remove
the
cabinet
rear
panel.4.
Remove
the
screw
and
lockwasher
securing
the
capacitor
box
cover
and
lift
off
thecover.5.
Tag
each
wire
before
removal,
then.
loosen
the
two
terminal
screws
(one
on
the
ter¬minal
board
and
one
on
the
potential
relay
Inside
the
capacitor
box)
for
the
fanmotor
cable.6.
Remove
the
two
screws
securing
the
armor
connector
cover
entering
the
capacitorbox.6-15

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Sorvall rc-5b