Updating Software - GE MAC 1600 Service Manual

Ecg analysis system
Hide thumbs Also See for MAC 1600:
Table of Contents

Advertisement

Maintenance
Electrical Safety Checks (cont'd.)
Step
1
AC power cord
ground prong
to exposed
metal surface
1
NC = Normal Condition; SFC = Single Fault Condition; N/A = Not Applicable
2
UUT = Unit Under Test

Updating Software

Software updates are provided on an SD card. Perform a software update as
described in this section.
1.
Power up the system with the system connected to AC power.
2.
Insert the SD card with the software update in the SD card slot in the rear panel
as shown in the following illustration.
The gold contacts are face up.
3.
From the Main Menu, press F5 (System Configuration).
104
Polarity
Condition
N/A
(ground lug)
MAC™ 1600
UUT
2
- ON
1
______ Ω
Pass/Fail
Result
Leakage
Current
Limits
Less than
200mΩ
2028451-183E

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sde

Table of Contents