Electromagnetic Compatibility
Guidance And Manufacturer's Declaration - Electromagnetic
Immunity
This system is intended for use in the electromagnetic environment specified in the
following table. It is the responsibility of the customer or user to assure that this
system is used in such an environment.
Immunity Test
Conducted RF
IEC
60601-1-2:2007
• IEC
61000-4-6:
2006
Radiated RF
• IEC
60601-1-2:2007
• IEC
61000-4-3:
2006
136
Compliance
Compliance
Test Level
Level
3 Vrms
Multilink
Lead set:
150 KHz to
80 MHz
3.0 Vrms
@ 2 Hz mod.
Value Lead
set:
3.0 Vrms
KISS Lead
set:
0.3 Vrms
Multilink
3 V/m
Lead set:
80 MHz to
2.5 GHz
3.0 V/m
@ 2 Hz mod.
Value Lead
set:
3.0 V/m
KISS Lead
set:
1.0 V/m
MAC™ 1600
Electromagnetic Environment –
Guidance
Portable and mobile RF
communications equipment should
not be used closer to any part of
the [equipment or system], including
cables, than the recommended
separation distance calculated fro the
equation applicable to the frequency
of the transmitter.
Recommended separation distance
Recommended separation distance:
Multilink Lead Set:
d = 1.2√P
Value Lead Set:
d = 1.2√P
KISS Lead Set:
d = 11.6√P
Recommended separation distance:
Multilink Lead Set:
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
Value Lead Set:
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
KISS Lead Set:
d = 3.5 √P 80 MHz to 800 MHz
d = 7 √P 800 MHz to 2.5 GHz