GE MAC 1600 Service Manual page 91

Ecg analysis system
Hide thumbs Also See for MAC 1600:
Table of Contents

Advertisement

6.
Replace the stainless steel tube with the notched end (1) on the side shown in
the following illustration.
7.
Reassemble the printer assembly to the main assembly as described in
"Replacing the Printer Assembly" on page
8.
Configure the Printhead Resistance as follows:
a.
b.
c.
d.
e.
f.
9.
Perform the applicable checkout procedures.
Refer to
2028451-183E
On the Main Menu, press F5 (System Configuration).
Press F6 (More) > F6 (More) > F5 (Service Setup).
When prompted, enter the Service Password.
If you do not know the password, contact GE Healthcare Technical
Support.
Select the Device Settings button.
In the Printhead Resistance field, type 1000
Press F6 (Save).
"Functional Checkout" on page
MAC™ 1600
83.
99.
Maintenance
91