Electrical Safety Checks (cont'd.)
Step
1
Forward
2
Neutral open,
Forward
3
Ground open,
Forward
4
Ground open,
Reverse
5
Neutral open,
Reverse
6
Reverse
1
Forward
2
Neutral open,
Forward
3
Ground open,
Forward
4
Ground open,
Reverse
5
Neutral open,
Reverse
6
Reverse
2028451-183E
Polarity
Condition
NC
Polarity
SFC
Polarity
SFC
Polarity
SFC
Polarity
SFC
Polarity
NC
Polarity
Patient Leakage Current to Ground
NC
Polarity
SFC
Polarity
SFC
Polarity
SFC
Polarity
SFC
Polarity
NC
Polarity
MAC™ 1600
UUT
2
- ON
1
______ µA
Pass/Fail
______ µA
Pass/Fail
______ µA
Pass/Fail
______ µA
Pass/Fail
______ µA
Pass/Fail
______ µA
Pass/Fail
______ µA
Pass/Fail
______ µA
Pass/Fail
______ µA
Pass/Fail
______ µA
Pass/Fail
______ µA
Pass/Fail
______ µA
Pass/Fail
Ground Continuity
Maintenance
Result
Leakage
Current
Limits
100 µA
300 µA
300 µA
300 µA
300 µA
100 µA
10 µA
50 µA
50 µA
50 µA
50 µA
10 µA
Resistance
103