Maintenance
6.
Lift the back of the base plastic (1) from the middle plastic (2), as shown in the
following illustration, until you can access the two cables that connect these
two assemblies.
7.
Disconnect the two cables from the mainboard.
8.
Set the base plastic assembly next to the middle plastic assembly as shown in
the following illustration and identify the power supply assembly (1).
9.
Using a small needle-nose pliers, disconnect pink ground connector from
AC outlet.
10.
Lift the power supply assembly (A in the previous figure) from the rails.
11.
Remove the power supply shield.
12.
Disconnect the white connector from the power supply circuit board by pinching
the white tab while pulling the connector.
13.
Disconnect the grounding cable from the back of the AC power connector.
14.
Remove the power supply assembly.
74
MAC™ 1600
2028451-183E