Maintenance
4.
Remove the two screws (1) shown in the following illustration.
5.
Lift the acquisition module assembly until you can access the black harness
connector on the other side of the acquisition module.
6.
If the system is equipped with the optional KISS pump, disconnect the gray
hose from the luer fitting.
7.
Remove the acquisition board assembly.
8.
Reassemble the acquisition board assembly. If the system does not have the
KISS option, go to step 9. If the system has the KISS option, go to step 10.
Use the following illustration for reference.
9.
Reassemble acquisition board assembly (systems with no KISS option).
a.
b.
76
Place the label (item 36) in the recessed hole of the plastic bezel.
Reverse the disassembly steps to reassemble.
MAC™ 1600
2028451-183E