8. Acquiring i-CEPH Images (Optional)
7.
You can see the patient through the Touch Screen. Make sure that the horizontal lines
displayed in the screen are parallel to the patient's Frankfurt line by using the Chinrest
UP/DOWN button.
8.
Let the patient maintain the posture as follows:
Close the mouth.
Close the eyes.
9.
Click NEXT button when the patient positioning is completed. The Rotating Unit will
come down to the exposure location. Then, READY button will be activated on the
Touch Screen and the Console Software.
10.
Press READY button. No X-ray will be emitted at this point.
<Console Software>
11.
Go to 8.5 X-ray Exposure to start the exposure.
90
Before conduct the next step, be sure to let the patient know that the
Rotating Unit will be lower to the exposure location and instruct him or
her to relax down.
When the equipment is not operable while the Rotating Unit is in the
exposure location, detach the Headrest from the equipment (detach
the Chinrest as well when it's attached) and open the Handle Frame to
bring out the patient immediately.
<Touch Screen>
PaX-i3D Green Premium™ User Manual