Technical Specifications - Vatech Premium PAX-i3D User Manual

Table of Contents

Advertisement

15.2

Technical Specifications

15.2.1
X-ray Generator Specifications
A-type: Superior Tube Option
High Voltage
Generator
X-ray Tube
PCT-90LH User Manual
Item
Model
Rated output power
Type
kVp
Normal/
Pulse
mA
Cooling
Total filtration
Added filtration
Manufacturer
Model
Focal spot size
Target Angle
Inherent Filtration
Anode Heat
Content
Duty Cycle
15. Technical Specifications
Description
DG-08C22S1
1.2 kW
Inverter
60 ~ 120 kV (1 kV increment ±10 %)
4 ~ 10 mA (1 mA increment ±20 %)
(PANO & i-CEPH: 0.5 mA increment,
CT: 0.1 mA increment ±10 %)
Cooling Protect
(Optional fan cooling ≥ 60 ℃)
Min. 2.5 mm Al
2.0 mm Al
Superior
SXR-130-15-0.5 (Stationary Anode type)
0.5 mm (IEC 60336)
15 degree
At least 1.1 mm Al equivalent at 80kV
21 kJ
1:20 or more
(Exposure time : Interval time)