Depending on the circumference of the patient's head, X-ray intensity
may be classified as Hard, Normal, or Soft :
Soft ≤ Normal ≤ Hard
5.
The values of tube voltage and current are configured automatically according to the
patient's gender / age group and X-ray intensity. Click Arrow button to zoom in. The
dose is adjustable by the ±1 kVp and ±0.5 mA respectively.
6.
When the exposure parameter setting is completed, guide the patient to the equipment
for positioning the patient. (Go to 8.3 Getting Prepared for Patient Positioning.)
PCT-90LH User Manual
On the Console Software, do not click NEXT button before
completing patient positioning.
On the Touch Screen, it's ok to press NEXT button after competing
parameter setting. (When NEXT button is pressed, a confirmation
window will appear.) But do not press OK button before completing
patient positioning.
8. Acquiring i-CEPH Images (Optional)