When you click NEXT (OK in Touch Screen) button,
 The Rotating Unit will come down to the positioning location.
 The vertical Laser Beam will be activated to make patient positioning
 The DAP (Dose Area Product), Scan Time and Exposure Time will be
7.
You can see the patient through the Touch Screen.
Make sure that the patient's Frankfurt line is tilted about 5˚ from the horizontal lines of
the screen.
8.
Let the patient maintain the posture as follows:
After swallowing once, open your mouth as wide as possible.
Close the eyes.
9.
Click NEXT button when the patient positioning is completed. The Rotating Unit will
come down to the exposure location. Then, READY button will be activated on the
Touch Screen and the Console Software.
10.
Press READY button. No X-ray will be emitted at this point.
<Console Software>
11.
Go to 7.5 X-ray Exposure to start the exposure.
PCT-90LH User Manual
easier.
displayed below Patient Information window.
Before conduct the next step, be sure to let the patient know that the
Rotating Unit will be lower to the exposure location and instruct him or
her to relax down.
When the equipment is not operable while the Rotating Unit is in the
exposure location, detach the Chinrest from the equipment and open
the Handle Frame to bring out the patient immediately.
7. Acquiring PANO Images
<Touch Screen>
73