Vatech Premium PAX-i3D User Manual page 133

Table of Contents

Advertisement

4.
Install the Headrest. (Only when Auto Ceph option selected.)
Put the Headrest into the Chinrest.
Put the Headrest Strap on the patient's head and turn the knob clockwise to hold
the head tightly.
5.
On the Console Software, Click NEXT button when the exposure parameter setting is
completed. On the Touch Screen, a confirmation window will appear after NEXT button
is clicked. Click OK button if the setting is correct.
(In case NEXT button has been already clicked, you will see pop-up window directly.)
<Console Software>
When you click NEXT (OK in Touch Screen) button,
 The Rotating Unit will come down to the positioning location.
 The vertical Laser Beam will be activated to make patient positioning
 The DAP (Dose Area Product), Scan Time and Exposure Time will be
6.
You can see the patient through the Touch Screen. Make sure that the horizontal lines
displayed in the screen are parallel to the patient's Frankfurt line by using the Chinrest
UP/DOWN button.
PCT-90LH User Manual
It is not mandatory to install the Headrest for CT mode. But it is highly
recommended to acquire the images with higher quality.
For more information about using the Headrest, refer to 8.3 Patient
Positioning for i-CEPH mode.
easier.
displayed below Patient Information window.
10. Acquiring ENT CT Images
<Touch Screen>
127

Advertisement

Table of Contents
loading

This manual is also suitable for:

Premium pct-90ln

Table of Contents