Vatech Premium PAX-i3D User Manual page 174

Table of Contents

Advertisement

16. Appendices
16.2.2
Leakage Dose
X-ray Leakage Dose Test is performed to protect patients against excessive and
unnecessary radiation that is not purposed and this document evaluates leakage dose
amount based on following standard defined by IEC regulation and has been
performed by covering each collimator region in use.
Standard
National
Deviation
International
Standard
IEC 60601-1-3
Test Condition
Test mode
PANO / Adult
Dental CT (21 x 19)
ENT CT (21 x 19)
i-CEPH
Measurement Overview
168
Terminology
limits leakage at 1M from the source to
Leakage
100 mR in 1hr
kVp
90
110
110
102
Permissive Range
mA
9.6
6.5
6.5
4.5
PaX-i3D Green Premium™ User Manual
sec
13.5
18.0
18.0
18.0