Insert the Bite Block into the Chinrest and cover it with a sanitary vinyl cover.
5.
Let the patient bite the Bite Block along its grooves with his/her front teeth.
6.
On the Console Software, Click NEXT button when the exposure parameter setting is
completed. On the Touch Screen, a confirmation window will appear after NEXT button
is clicked. Click OK button if the setting is correct.
(In case NEXT button has been already clicked, you will see pop-up window directly.)
<Console Software>
When you click NEXT (OK in Touch Screen) button,
The Rotating Unit will come down to the positioning location.
PCT-90LH User Manual
The sanitary vinyl cover is for single use only. It should be replaced
for each patient. Be sure to use the approved vinyl cover.
Clean the Chinrest and the Bite Block with ethanol and wipe with a
dry towel before the next patient.
7. Acquiring PANO Images
<Touch Screen>