Guidance and manufacturer's declaration - electromagnetic immunity
The PCT-90LH is intended for use in the electromagnetic environment specified
below. The customer or the user of the PCT-90LH should assure that it is used in
such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1) At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects, and
people.
PCT-90LH User Manual
IEC 60601-
Compliance
1-2
level
Test level
3 Vrms
V
=3Vrms
150 kHz to
1
80MHz
3 V/m
80 MHz to
E
=3V/m
1
2.5GHz
16. Appendices
Electromagnetic environment -
Guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the PCT-90LH, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance:
where P is the maximum output
power rating of the transmitter in
watts(W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters(m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance level in each
b
frequency range.
Interference may occur in the
vicinity of the equipment marked
with the following symbol :
a